GradyVet 7903B Infusion Pump

The new GradyVet 7903B infusion pump is one of our veterinarian favorites! It is compatible with any standard IV set and stores up to 200 editable brands. The pump has 5 infusion modes with rates from 0.1ml/hr to 2000ml/hr. WiFi capability allows for optional connection to a central monitor.
  1. 4.3 inch color touch screen display
  2. Easy calibration
  3. Quiet operation
  4. Vertical Line placement
  5. Compact design with pop-up handle
  6. Drug library
  7. Run Memory (patient history log)
  8. Available in pole mount or cage holder style
  9. Rapid infusion start based on last treatment
  10. 10 hours battery back-up time
  11. IP44 waterproof
  1. Standard IV sets
  2. Small size 1.45 kg / 3.19 lbs.
  3. Flow rate .01 – 2000 ml/hr
  4. Bolus up to 2,000 ml/h
  5. Battery up to 5 hrs. running at 25 ml/h
  6. KVO at 0.01 ml/h – 5 ml/h
  7. VTBI 0 – 9999 ml
  8. Up to 2,000 ml per hour delivery rate
  9. Battery up to 5hrs at 25 ml/h
Specification Parameters
Infusion apparatus specification
10-60 drips/ml
Infusion Rate
0.1-2000ml/h, minimum step is 0.01ml/h
<100ml/h step is 0.01ml/h,
<1000ml/h step is 0.1ml/h,
≥1000ml/h step is 1ml/h
Infusion accuracy
≤±5%
Drop rate
Infusion set drip pot setting range 10—60 drips/ml, drip rate 1-2000 drops/min, step is 1 drops/min
Drop rate accuracy
≤±5%
Bolus Rate (Bolus)
1-2000ml/h, minimum step is 0.01ml/h
<100ml/h step is 0.01ml/h,
<1000ml/h step is 0.1ml/h,
≥1000ml/h step is 1ml/h
Bolus Rate accuracy
≤±10%
Purge rate
100-2000ml/h, minimum step is 0.1ml/h
<1000ml/h step is 0.1ml/h,
≥1000ml/h step is 1ml/h
Purge rate accuracy
≤±5%
Mirco Mode
100ml/h to 1200m/l, programmable
VTBI
0-9999.99ml, minimum step is 0.01ml
Infusion accuracy
≤±5%
Total Volume Infused
0-9999.99ml
KVO rate
0~5ml/h, minimum step is 0.01ml/h
KVO rate accuracy
≤±10%
Micro mode setting range
0.1~200ml/h
Time Range
1s-99hrs59min59s
Accumulated bubbles
50~1000ul/15min
Acti agentia
0.0199999
Volume
0.01
Conc.
0.0199999
Dose rate
0.019999
Cumulative Bubble
50~1000μl 15
Single fault bolus Volume
≤ 2ml
Anti-bolus volume
≤ 0.2ml
Fuse Type
T2AL 250V
Dimensions
131.5*90*138mm (No fastening clamp, no drop sensor hook)
Weight
single battery with clamp≤1.45kg double batteries with clamp ≤1.55kg
Power Supply
AC power supply
100V-240V AC,50Hz/60Hz,0.25A-0.1A
Input power
50VA
DC power supply
DC 10V-16V,1.5A-0.94A
Battery
Battery quantity
1piece (standard) or 2 pieces (optional)
Battery type
Lithium battery
Rated battery voltage
7.4V
Battery capacity
2600mAh (1 piece battery) or 5200mAh (2 pieces batteries)
Charging time
Off status≤4h(1 piece battery) or ≤8h((2 pieces batteries)
Running time
Use a new battery full of electricity to power: standard (1pc)battery, running at 25 ml/h, the running time from start-up to exhaustion of alarm batteries is not less than 5 hours.

optional(2pcs)battery, running at 25 ml/h, the running time from start-up to exhaustion of alarm batteries is not less than 10 hours.

standard(1pc)battery, running at 2000 ml/h, the running time from start-up to exhaustion of alarm batteries is not less than 2.5hours. optional(2pcs)battery, Running at 2000 ml/h, the running time from start-up to exhaustion of alarm batteries is not less than 5 hours.
Alarm
Alarm signal sound pressure level
When the sound is set at lowest level, alarm signal sound pressure level ≥50dB(A)

When the sound is set at highest level, alarm signal sound pressure level ≤80dB(A)
Alarm information
VTBI near end, VTBI infused, Pressure high, Battery nearly empty, Battery empty, system error, No power supply, Reminder alarm, Standby time expired, KVO finished, Drop sensor connection, Drop error, Empty bottle, Single bubble, Cumulative Bubble, Door Open, Occlusion pre-alarm, Drop in pressure, Drug dose limits exceeded, Backup battery power exhaustion